Device establishmentActive, RegisteredOfficial Action Indicated

Microspecialties LLC

Microspecialties LLC is an FDA-registered device establishment (FEI 3002616700) located in Middletown, Connecticut. The facility is active, registered. Its most recent FDA inspection on record (2025-02-18) was classified OAI.

Recalls (1)

Jun 5, 2024 Class II
Could not confirm products were exposed to the proper sterilization cycle can potentially cause post operative wound infection or injection site infection.

Recall data from openFDA enforcement reports (FDA Recall Enterprise System).

Establishment record

FEI Number
3002616700
Address
430 Smith St, Middletown, Connecticut
Establishment type
Device establishment
Products / operations
Ophthalmic, Repack or Relabel Medical Device, Manufacture Medical Device, Unknown, General, Plastic Surgery, Complaint File Establishment per 21 CFR 820.198, Manufacture Medical Device for Another Party (Contract Manufacturer), Orthopedic, Manufacture Device in the United States for Export Only, General Hospital
Registration status
Active, Registered

Verify this record in FDA source data ›

Inspection history

Inspections on file
Feb 18, 2025 · Official Action Indicated (OAI)
Feb 18, 2025 · No Action Indicated (NAI)
Sep 4, 2014 · Voluntary Action Indicated (VAI)
Feb 3, 2012 · Voluntary Action Indicated (VAI)
Feb 3, 2012 · No Action Indicated (NAI)
May 10, 2011 · Official Action Indicated (OAI)
Form 483 observations
Feb 18, 2025, 5 observations: DMR - not or inadequately maintained; Nonconforming product control; Lack of or inadequate process validation
Sep 4, 2014, 3 observations: Environmental control Lack of or inadequate procedures; Lack of or inadequate complaint procedures; Purchasing controls, Lack of or inadequate procedures
Feb 3, 2012, 4 observations: Documentation of validated process performance; Documentation; Process control procedures, Lack of or inadequate procedures
May 10, 2011, 7 observations: Lack/Inad procedure-Monitoring/Control of Validated Proces; Lack of or inadequate In-process acceptance procedures; Lack of procedures, or not maintained

FDA record timeline

FDA inspection
OAI classification
FDA inspection
NAI classification
Form 483
5 observations
Recall
Class II — Could not confirm products were exposed to the proper sterilization cycle can potentially cause post operative wound infection or injection site infection.
Source ›
FDA inspection
VAI classification
Form 483
3 observations
FDA inspection
VAI classification
FDA inspection
NAI classification
Form 483
4 observations
FDA inspection
OAI classification
Form 483
7 observations

Combined chronology of the public records shown on this page. FDA coverage varies by dataset and year.

Questions

Is Microspecialties LLC registered with the FDA?

Yes. Microspecialties LLC is an FDA-registered device establishment (FEI 3002616700) in Middletown, Connecticut. Its current registration status is Active, Registered.

Has Microspecialties LLC received an FDA warning letter?

FDA has not posted a warning letter for Microspecialties LLC. FDA publishes only a subset of enforcement records, so this is not a guarantee.

What does Microspecialties LLC do?

Microspecialties LLC's FDA-registered operations are: Ophthalmic, Repack or Relabel Medical Device, Manufacture Medical Device, Unknown, General, Plastic Surgery, Complaint File Establishment per 21 CFR 820.198, Manufacture Medical Device for Another Party (Contract Manufacturer), Orthopedic, Manufacture Device in the United States for Export Only, General Hospital.

Related facilities

Source: FDA public data · status as of 2026-10-12

The Quality Brief